Zeroing bird power sensors, Figure 38 zeroing, message 1 – Bird Technologies SA-6000XT-Manual User Manual
Page 62

46
Zeroing Bird Power Sensors
Over time, the sensor’s “zero value” (reading with no applied RF power) can
drift, making all readings inaccurate by this value.
Example - If the zero value is –2 μW, measuring a 5 mW
signal will give a reading of 4.998 mW, a 0.04% error. Peri-
odically re-zero the sensor.
Note: This applies to TPS and WPS power sensors.
Note: For best results, make sure the sensor has been connected to
the Site Analyzer and the SA turned on for at least 5 minutes. Do not
apply RF power to the sensor during calibration.
Figure 38 Zeroing, Message 1
CAUTION
When using terminating power sensors do not exceed 2 W average or 125 W
peak power for 5 μs. Doing so will render the sensor inoperative.
Press from
Measure
Power Mode
Press to
begin the
calibration