beautypg.com

Figure 13 setting match mode scale limits, Setting the measurement data points – Bird Technologies SA-6000XT-Manual User Manual

Page 33

background image

17

Figure 13 Setting Match Mode Scale Limits

Setting the Measurement Data Points

Select the number of data points to take during a measurement. There are three
data point options, 238 points, 475 points, and 949 points. Increasing the number
of data points increases the measurable distance in distance-to-fault measure-

ments and increases the detail in Measure Match mode measurements. When
saving a trace, the number of data points collected (238, 475, or 949) are saved.

The default number of data points is 238. Selecting 475 data points provides
twice as many measurement points as 238, but it takes approximately twice as
long for the trace to sweep and display. Selecting 949 data points provides four
times as many measurement points as 238, but it takes approximately four
times as long for the trace to sweep and display.
A progress bar, just below the x-axis, moves from left to right indicating the
progress of the sweep. When the progress bar reaches the right side of the
screen, the sweep is finished.
When the number of measurement data points are changed, the current trace
is erased and the Save softkey is disabled until the sweep finishes and the
screen displays the first new trace. If the Save softkey is pressed before the first
sweep has completed, the Save softkey icon (the red bar) will blink until the
sweep finishes and then the display will change to the Save and Recall screen (in

response to pressing the Save softkey).
1.

Enable the Measure Match Mode. See “Selecting the Measure Match
Mode” on page 13.

2.

Set the frequency. See “Setting the Measurement Frequency” on page 14.

3.

Calibrate the Site Analyzer. See “Calibration” on page 11.

4.

Press the Config key.

5.

Press the Scale softkey.

6.

Press the left or right scroll key to select Pts.

7.

Press the up scroll key to open the selection list.

8.

Press the up or down scroll key to select a data point option (238, 475, 949).

9.

Press the Enter key to accept the selection and return to the configuration screen.

This manual is related to the following products: