Figure 32 setting a limit line – Bird Technologies SA-6000XT-Manual User Manual

Page 54

background image

38

1.

Enable either the Measure Match Mode, see “Selecting the Measure Match
Mode” on page 13.

2.

Set the frequency. See “Setting the Measurement Frequency” on page 14.

3.

Calibrate the Site Analyzer. See “Calibration” on page 11.

4.

Enable the Fault Location Mode, if necessary. See “Selecting Fault Location
Mode” on page 26.

5.

Press the Limit Test key to activate the limit line:

a.

Test ON Aud (line and sound)

b.

Test ON (line only)

c.

Test OFF
Note: A limit line value can be set with the limit test turned OFF, but
will not see the line displayed.

6.

Press the Config key.

7.

Press the Limit Line softkey.

8.

Use the keyboard to enter a limit value then press the Enter key.

9.

Press the Esc key to exit the configuration screen.

Figure 32 Setting a Limit Line