Chapter 2 calibration, Calibration accessories, Calibrating the site analyzer – Bird Technologies SA-6000XT-Manual User Manual
Page 27
11
Chapter 2
Calibration
Calibration Accessories
Calibration Combination (Cal Combo) which contains:
One 50 ohm load
One Open standard
One Short standard
Note: When using a test cable connected to the Site Analyzer, attach
the Cal Combo to the end of the cable during calibration.
Test Cable (optional) – A phase-stable cable will provide consistent,
reliable results.
Note: When using an extension cable, a phase stable cable is needed
to ensure performance and accuracy. Bird offers optional phase stable
cables that will not degrade the performance of the Site Analyzer.
Using a non-Bird phase stable cable may decrease and degrade the
accuracy of the system.
Calibrating the Site Analyzer
For best results, set the frequency and calibrate the Bird Site Analyzer immedi-
ately before taking measurements.
Note: When calibrated, the Site Analyzer’s main screen will display
“Calibration: FULL”. When uncalibrated, the Site Analyzer’s main
screen will display “Calibration: OFF”.
Note: Calibration is performed at the highest number of data points
(949 points) and the calibration graphs are displayed using the cur-
rently selected display method (envelope, maximum, or minimum).
1.
Press the Calibrate key.
2.
Connect the Cal Combo “Open” connector to the unit antenna port.
3.
Press the Open softkey then wait for the unit to beep and display Done.
4.
Connect the Cal Combo “Short” connector to the unit antenna port.
5.
Press the Short softkey then wait for the unit to beep and display Done.
6.
Connect the Cal Combo “Load” connector to the unit antenna port.
7.
Press the Load softkey then wait for the unit to beep and display Done.
8.
Remove the Cal Combo from the unit.