beautypg.com

Sensaphone IMS-4000 Users manual User Manual

Page 47

background image

45

Chapter 2: IMS-4000 Software

Calibration: Enter a positive or negative offset in this field to calibrate the sensor value.
High Limit: Enter the sensor high alarm limit in this field. When the value exceeds this limit for the duration of the recogni-

tion time, and if the channel is enabled, an alarm will be tripped.
Low Limit: Enter the sensor low alarm limit in this field. When the value falls below this limit for the duration of the recog-

nition time, and if the channel is enabled, an alarm will be tripped.
Alarm Class: Click the drop down arrow and select the appropriate class for this sensor/channel. If you prefer you can create

your own classes on the Class Setup screen.

Figure 19: Alarm Class dropdown menu

Custom Voice: Click the drop down arrow and select the custom voice message you would like assigned to this sensor/chan-

nel. Voice messages can be recorded on your PC and uploaded into the IMS-4000 on the Custom Voice Manager screen (See

Recording and Uploading Voice Messages).

Figure 20: Custom Voice dropdown menu

Wait (Recognition) Time: This is the time required for a fault condition to qualify as an alarm event. The sensor/channel

must remain beyond the limits or in a fault condition continuously for this entire period of time in order to become an

alarm.
Reset Time: This is the time allowed for an acknowledged alarm’s fault condition to be corrected before the IMS-4000 resets

(reactivates) the alarm and begins the message delivery process all over again. The minimum reset time is 30 minutes.
Alarm on Return to Normal: This feature makes the unit send a message (or messages) when a condition has moved within

its programmed alarm limits or has gone from an alarm state to a Normal state. The person who acknowledged the alarm—

in addition to all Inform Only contacts in the alarm class—will receive the message.
SNMP Trap Level: When sending alarm messages via SNMP, this sets the priority of the SNMP trap. The choices are Major

or Minor.

alarM response Via The powerGaTe, powerGaTe2, or CaMera

powerGaTe

When an alarm occurs, you can have PowerGate or PowerGate2 outlets automatically turn ON, turn OFF, or CYCLE power

to a device. Cycling will switch an outlet OFF for 10 seconds and then switch it back ON. To set up a PowerGate, click the

Alarm Response button on the Channel Setup Screen. A sample Alarm Response screen is shown below.

Figure 21: Alarm Response screen for PowerGate

Select Response Type: Choose “PowerGate.”