The data is ordered according to its performance schedule. If a
measurement was carried out using the Mode from Database, then next to
the measurement date and time the record displays mode name. in order
to print data place the cursor (black triangle) next to a needed record and
press Setup key.


The detailed data on a measurement is displayed (data order as on a
printout).


In order to print data press PRINT key. The set of data is sent via RS 232
interface to a connected printer / computer.

19.

MEANS OF USING A MOISTURE ANALYZER FOR DRYING IN
TEMPERATURE OVER 160

°C

In case of drying a sample in a temperature ranging between 161

°C – 250°C the

maximum time of maintaining the temperature is determined proportionally from the
range: 1 hour for 161

°C – 20 minutes for 250°C.

When drying a sample in temperature 250

°C, the maximum temperature value is

maintained by the instrument for 20 minutes, and then the software decreases the
temperature value to 160

°C without stopping the drying process.

Time interval of decreasing the drying temperature from 250

°C to 160 °C is 20

minutes.

In case of drying mode „Quick“ the maximum drying temperature is exceeded by 30
% but its value is not higher than the temperature value set in instrument’s factory
menu.

In case of drying mode „Step“ the maximum time interval of maintaining a set
temperature value in each step is limited to 20 minutes.