User adjustment – RADWAG MAX 60 Moisture Analyzer User Manual

Page 24

background image

- 24 -

The moisture analyzer gives a command on
determining mass of an empty pan, followed
by a command to place a 50 g weight
(standard) on the pan.

a) After completing adjustment process the moisture analyzer returns to

submenu P1 - Adjustment

b) Press ESC for a few times until the display indicates below message

c) Depending on needs press ENTER or ESC key.
d) The moisture analyzer returns to the main window (weighing mode).

8.1.2.

User adjustment

User adjustment should be carried out with a mass standard (weight) in
accuracy class F

2

.

Enter group of parameters P1 Adjustment.
Set the marker next to a parameter 02 User
adjustment.

Press RIGHT ARROW KEY. The moisture
analyzer displays a command to enter mass
of an adjustment weight to its memory.
in the displayed mass value the first digit is
flashing, i.e. it can be modified.

Use navigating arrow keys to enter the new mass value of the external adjustment
weight.

Accept entered mass of the adjustment
weight. Then the moisture analyzer
proceeds to the adjustment process and
displays corresponding mesages on its
display.