beautypg.com

Cooperation with a printer or a computer, Connecting cables – RADWAG MAX 60 Moisture Analyzer User Manual

Page 77

background image

- 77 -

14.

COOPERATION WITH A PRINTER OR A COMPUTER

Each pressing of < PRINT > key causes sending to a connected printer
or computer a signal corresponding to current mass indication (display
status) with enabled measuring unit. The default baud rate setting of a
balance is 9600 bit/sec. If a peripheral device (a printer or a computer)
requires other baud rate settings then it needs to be changed in the
moisture analyzer’s menu (see point 13 of this user manual).

14.1.

Connecting cables

A balance: slot DB 9/F – a KAFKA printer: slot WM 560

3 (TxD)

1 (RxD)

5 (GND)

3 (GND)

7- 8 shorted


A balance: slot DB 9/F – a computer: slot DB 9/F (if handshake is disabled)

2 (RxD)

3 (TxD)

3 (TxD)

2 (RxD)

5 (GND)

5 (GND)

4 - 6 shorted

4 - 6 shorted

7- 8 shorted

7 - 8 shorted


A balance: slot DB 9/F – a computer: slot DB 9/F (if handshake is enabled)

2 (RxD)

3 (TxD)

3 (TxD)

2 (RxD)

4 (DTR)

4

5 (GND)

5 (GND)

6 (Tara)

6

7 (RTS)

7 (CTS)

8 (CTS)

8 (RTS)

9 (Print)

9


A balance: slot DB 9/F – a computer: slot DB 25/F

2 (RxD)

2 (TxD)

3 (TxD)

3 (RxD)

5 (GND)

7 (GND)

7 - 8 shorted

4 - 5 shorted

6 - 20 shorted


This manual is related to the following products: