RADWAG MAX 60 Moisture Analyzer User Manual
Page 30

- 30 -
Press RIGHT ARROW navigating key. The
moisture analyzer opens a window for
entering the temperature value for carrying
out the drying test. Use the navigsting keys
to enter the desired temperature value and
press ENTER button to accept.
The moisture analyzer displays another
message for entering the permissible
temperature error that is acceptable by an
operator. Use the navigating keys to enter
the desired error size and press ENTER
button to accept.
The moisture analyzer displays another
message for entering the factory number of
the adjustment set used for carrying out the
test. Use the navigating keys to enter the
factory number data and press ENTER
button to accept.
After accepting the factory number of the
adjustment set, the software automatically
starts heating the drying chamber until
reaching a pre-defined temperature value.
The moisture analyzer’s display indicates
data on tim and temperature of the drying
chamber sensor.
The temperature is maintained for 8 minutes
(as in case of temperature adjustment).
After this time interval, a message box is
opened on moisture analyzer’s display for
entering the temperature value as recorded
by the control thermometer set assembled
inside the drying chamber. Use the
navigating keys to enter the temperature
value from the control thermometer and
press ENTER button to accept it.
Then the display opens a mssage box with
the temperture error of the drying chamber
as recorded by the control thermometer set.