User menu, Moisture analyzer adjustment, Adjustment – RADWAG MAX 60 Moisture Analyzer User Manual
Page 23

- 23 -
8.
USER MENU
User menu is accessible on pressing SETUP key on moisture analyzer’s overlay.
The user enables: moisture analyzer adjustment, setting transmission parameters,
setting time and date and their format, and also other user parameters. This section
of the user manual presents software possibilities of the moisture analyzer and
means of changing the parameter settings.
8.1.
Moisture analyzer adjustment
8.1.1.
Adjustment
Should a moisture analyzer MAX series be used as a laboratory balance, then
before carrying out the measurements it has to be adjusted suing and external
mass standard / weight).
Adjustment is a set of processes determining relation between the value
indicated by a moisture analyzer / balance, and mass of a weighed standard
(adjustment weight). The process enables correcting the indication if
necessary.
Adjustment process on a moisture analyzer MAX series should be carried out
with use of a mass standard / adjustment weight class F
2
. The weight is an
additional equipment of the moisture analyzer.
Moisture analyzer type
Adjustment weight mass
MAX
50g
Enter group of parameters P1 Adjustment.
Set the cursor next to a parameter 01
External adjustment.
Press RIGHT ARROW KEY, which displays
a command to unload the moisture
analyzer’s pan (the pan has to be empty).
After emptying the pan press ENTER key.