Temperature sensor cleaning – RADWAG MAX 60 Moisture Analyzer User Manual

Page 82

background image

- 82 -

After cleaning, thoroughly dry each of moisture analyzer’s components.
Do not allow any contamination or liquid to fall into the moisture analyzer.

Assembly cleaned and dried components in accordance with above scheme.

17.2.

Temperature sensor cleaning

In order to ensure correct temperature measurements during testing procedure, pay
special attention to cleanness of the temperature sensor (1).
Any cleaning activities of the temperature sensor should be carried out while
maintaining special precautions.
Temperature sensor should cleaned using a soft fabric with addition of a mild
detergents. Do not use any abrasive agents or solvents, as they may damage the
temperature sensor.


While cleaning make sure not to touch heating filament, as it may damage it.

If necessary it is permissible to clean filament’s shield. Clean the filament’s shield
using a soft and dry fabric only. Do not touch the filament while cleaning the shield.