beautypg.com

Auto switch off settings – RADWAG MAX 60 Moisture Analyzer User Manual

Page 61

background image

- 61 -

10.5.

Auto switch off settings

Finishing of a drying mode can be carried out as:
-

Automatic 1 (change ratio 1mg/10s)

-

Automatic 2 (change ratio 1mg/25s)

-

Automatic 3 (change ratio 1mg/60s)

-

Automatic 4 (change ratio 1mg/90s)

-

Automatic 5 (change ratio 1mg/120s)

-

Time defined (max. time value 99h : 59min.)

-

Manual (on pressing a button)

-

Custom (a user should specify change in mass

∆m 0,1-9,9mg over time ∆t

max 2,55sec.

-

Custom 2 (a user should specify change of humidity value

∆%M over time

60s)

-

Test (intended to select the finish-mode criterion for a tested sample)

Procedure:

Go to menu. Use DOWN navigating arrow
key to place the cursor (black triangle)
next to a parameter Finish mode.

The criteria on available settings are described in point ref. to “shortened drying
menu” (see point 9.2).

10.6.

Setting the type of displayed and printed measurement result

Depending on user needs, the moisture analyzer MAX series enables setting
the type of displayed and printed result of a drying process. The user can
select between four available settings: %M, %D, %R, g.

Procedure:

Go to software menu. Use DOWN
navigating arrow key to place the cursor
(black triangle) next to a parameter
Result.

If option “Chart” is enabled, then the printed result is %M.

This manual is related to the following products: