Press right arrow navigating key to
enable selection of a drying profile (name
of a drying mode is previewed above the
mode’s graphic presentation. The
pictogram denoting drying mode
selection menu is flashing in the upper
left corner of the display)

Press UP or DOWN navigating arrow
keys to select one of available drying
modes: Standard, Quick, Mild or Step.

If selected drying mode is QUICK, then
set the drying temperature and heating
time in temperature exceeded by 30 %
than the actual drying temperature. The
heating time is calculated from the
moment of reaching the set temperature
by the moisture analyzer.
If selected drying mode is STEP, then set
the drying temperatures for thresholds 1
and 2 and the drying temperature and
their corresponding time intervals for
each of the thresholds. The heating time
in each of the thresholds is calculated
from the moment of reaching a set
temperature in each of the steps.

On selecting appropriate drying mode
press the right arrow navigating key to go
to the following step of drying mode