- 26 -

1. Thermometer
2. Thermometer holder with a shield

Adjustment is a simple process aimed at adjusting the temperature sensor
assembled in moisture analyzer’s drying chamber. Before carrying out the
temperature adjustment process, install the adjustment set inside the drying
chamber following below figures.

Step 1.

Remove from the drying
chamber the following:

Disposable pan

Pan holder

Pan bracket

Chamber cover