beautypg.com

Moisture analyzer temperature stabilization period, Moisture analyzer’s menu – RADWAG MAX 60 Moisture Analyzer User Manual

Page 18

background image

- 18 -

On/Off key – switch on/off display of moisture analyzer. On
switching off the display of the device, other subassemblies
are continuously supplied, and the moisture analyzer is in
standby mode.

Display key – changes type of data displayed during and
after drying process.

Setup key – enter to main menu.

Test menu key – function key for selecting drying modes.

5.1.

Moisture analyzer temperature stabilization period

Before start of measuring processes, it is necessary to wait until the moisture
analyzer is thermally stabilized. It is a period of so called moisture analyzer’s
self-heating.
For moisture analyzer MAX series which before plugging to mains was stored
in room temperature, self-heating period takes approximately 30 minutes. For
moisture analyzers that were stored in much lower temperatures before
plugging to mains (e.g. during winter period) thermal stabilization should last
approximately 4 hours.
During self-heating period the indications on moisture analyzer’s display may
change. Correct operation of a moisture analyzer is possible within
temperature range specified in the technical parameters, see point 1.
It is recommended that ambient temperature changes at moisture analyzer’s
place of use are very small (slow).

6.

MOISTURE ANALYZER’S MENU

User menu of the moisture analyzer MAX series is divided into two sections.
The first one is made available on pressing SETUP key. This menu enables
setting the following parameters: moisture analyzer and temperature sensor
adjustment, setting date and time format, programming printouts and
communication parameters with peripheral equipment, and setting user profile
parameters.
The second menu is accessible on pressing TEST MENU key, and it is
dedicated to setting parameters of basic drying mode, including: drying
temperature, drying mode, time and other parameters. The menu additionally
provides access to the library of drying programs. The library stores up to 99
different drying procedures containing detailed settings on drying modes of
various substances.

This manual is related to the following products: