8.6.5.

Enabling non-standard printout templates

Enabling a non standard printout template, e.g. printout no. 1:

Enter group of parameters P6 Printouts. In a parameter determining printout type
01 Printout no. enter 1: printout 1 and in parameter 02 printout 1 start enter the
text string of printout start. Available values for start of a non-standard printout: from
1 to 80 !

After entering the number of a text string starting a printout, set number of a

text string finishing the non-standard printout template:
Enter parameter settings 03 printout 1 stop and enter the text string of printout
stop (end). Available values for start of a non-standard printout: from 1 to 80!

After accepting the selected parameters and returning to a weighing mode with

procedure of saving changes each pressing of PRINT key causes printing of a
declared non-standard printout template.

8.6.6.

Non-standard printout templates in drying mode

The process of designing non-standard printout templates for drying mode is the
same as in case of weighing mode. The only feature of the printout template in
dying mode differentiating it from the printout in weighing mode is the need to
determine thee sections of a printout template (see standard printout in drying
mode): Header, Line and Footer.

An example of determining a printout template for a drying mode:

Header – Printout 2
Line – Printout 3
Footer – Printout 4

Printout 2