beautypg.com

RADWAG MAX 60 Moisture Analyzer User Manual

Page 79

background image

- 79 -

Function

GIVE SOFTWARE REVISION

Command

RV CR LF (sends software name)

Function

GIVE OR CHANGE DATE IN A BALANCE

Command

PD CR LF (causes sending data on set date or changes the date)

Function

GIVE OR CHANGE TIME IN A BALANCE

Command

PD CR LF (causes sending data on set time or changes the time)


Function

GIVE ENABLED WORKING MODE

Command

PM CR LF (causes sending by a balance a code of enabled
working mode)

Function

SEND SETUP

Command

PS CR LF (causes sending the complete setup - parameter
printout)

Function

“BEEP” SOUND

Command

B CR LF (causes immediate activating of beep sound in a
balance)


Function

GIVE LAST CODE ERROR

Command

ER CR LF (causes sending code of last saved error in a balance)

Function

DISABLE AUTOZERO

Command A0 CR LF (causes disabling (switching off) autozero operation)

Function

ENABLE AUTOZERO

Command A1 CR LF (causes enabling (switching on) autozero operation)

Caution!

If a non-existing or incorrect command finished with CR LF is sent to a
balance, it responses with E S CR LF. Spaces between characters should
be omitted, as they are added only for the purpose of proper legibility.