beautypg.com

Drying test – RADWAG MAX 60 Moisture Analyzer User Manual

Page 29

background image

- 29 -

P1 05 Report printout :

1 : YES – report is printed

P1 05 Report printout:

0 : NO – report is not printed

Remember that on enabling report printout (set to 1), the report is generated
and sent automatically.

An example of a printout of a report from adjustment process:

The printout content depends on

parameters set in group P2 GLP (see point
8.2
of this user manual).
Variables that are
printed have an attribute set to 1: yes.

Apart from the content set in this menu

group, a report contains the following data:
mass of an adjustment weight stored in
moisture analyzer memory from last
completed adjustment process (described as
Curr:), mass of an adjustment weight
determined in current adjustment process
(described as Adj: (or Cal) and a deviation
from adjustment process (described as
Dev:), i.e. the difference between the two
mass values.

8.1.6.

Drying test

Parameter enables monitoring correctness of temperature
indications as recorded by the moisture analyzer’s internal thermometer.
The testing procedure is carried out using a dedicated set (a thermometer with a
holder) that is available as additional equipment of a moisture analyzer MAX
series. (It is the same set that is used for carrying out temperature adjustment
process). Means of assembling the set is described in the chapter on
temperature adjustment of this user manual.

Enabling the testing procedure requires assembling the control set inside the
moisture analyzer’s drying chamber, and entering a menu group
in instrument’s parameters. Start the process by following below specified
instructions.
Remember to assembly the control thermometer set before starting the testing
procedure and then close the drying chamber lid.

Enter menu group P1 Adjustment.
Set the marker next to a parameter 06
Drying test

This manual is related to the following products: