17.4.

Troubleshooting

Problem: no reaction to pressing main switch on/off key (display remains dark).
Probable cause:

− No voltage in mains

− Damaged power cord

− Damaged fuse of the moisture analyzer
− Moisture analyzer defect


Problem: too long pending time for drying process finish.
Probable cause::

− Selected finish mode of drying process is incorrect. Carry out experimental

tests to select a proper means of finish mode.


Problem: measurements are not repeatable
Probable cause:

− non-uniform sample content. Prepare the sample from larger amount of

substance.

− the drying time is too short. Change testing process finish mode criteria.
− drying temperature is too high, causing sample’s oxidizing. Lower drying

temperature.

− tested sample boils. Lower the drying temperature.
− the temperature sensor is contaminated or defected. Clean the

temperature sensor.

− the workbench on which the moisture analyzer is located is unstable.

Change the location of instrument’s operation.

− - the ambient area is incompatible with the requirements (vibrations, drafts,

etc.). Change the ambient conditions for compatible ones according to the
guidelines of this user manual.