Troubleshooting – RADWAG MAX 60 Moisture Analyzer User Manual
Page 84
- 84 -
17.4.
Troubleshooting
Problem: no reaction to pressing main switch on/off key (display remains dark).
Probable cause:
− No voltage in mains
− Damaged power cord
− Damaged fuse of the moisture analyzer
− Moisture analyzer defect
Problem: too long pending time for drying process finish.
Probable cause::
− Selected finish mode of drying process is incorrect. Carry out experimental
tests to select a proper means of finish mode.
Problem: measurements are not repeatable
Probable cause:
− non-uniform sample content. Prepare the sample from larger amount of
substance.
− the drying time is too short. Change testing process finish mode criteria.
− drying temperature is too high, causing sample’s oxidizing. Lower drying
temperature.
− tested sample boils. Lower the drying temperature.
− the temperature sensor is contaminated or defected. Clean the
temperature sensor.
− the workbench on which the moisture analyzer is located is unstable.
Change the location of instrument’s operation.
− - the ambient area is incompatible with the requirements (vibrations, drafts,
etc.). Change the ambient conditions for compatible ones according to the
guidelines of this user manual.