22. USER ADJUSTMENT

An option only for non-verified scale

Scales require to recalculate internal divisions to more suitable ones (e.g.

g, kg etc.). In order to do this they require an adjustment factor. It is adjusted

during the adjustment procedure using a mass standard. Adjustment should

be carried out if weighing a standard mass shows a different mass value.

To enter < User Adjustment>,

press

key and then

:

User Adjustment”

.

22.1. Adjusting procedure

Enter submenu < User Adjustment > according to ch. 22

and select:

Adjustment”

,

After entering the parameter the following message box appears:

Take the load off the pan of platform 1,

Press button

. The following message appears during

adjusting start mass: „Evaluation of starting mass”,

After the procedure has been completed the following message

box appears: