beautypg.com

RADWAG WLY Scales User Manual

Page 103

background image

103

Press

(estimate piece mass), then the editing field is displayed

together with the screen keyboard,

Enter a value and confirm it by pressing

, then the following

message is displayed: (where xx – the value

entered before),

Put the declared quantity of pieces on the pan and when the result

is stable (symbol

) confirm it by pressing

.

The program automatically calculates reference unit mass and

causes starting <

Parts counting >

.


Notice:

The total mass of all pieces put on the weight pan cannot be greater

than the weighing range;

The total mass of all pieves put on the weight pan cannot be less than

the value declared in parameter „Minimum reference mass” (see
section 26.2.2). If the condition above is not fulfilled the scale dispalays
the message ;

The mass of one unit may not be less than 0.1 of reading division

of the scale. If this condition is not fulfilled the scale dispalays the
message: .

26.5. Setting the reference mass by entering single piece mass directly

to the database

After selecting a product from the product database a mass of single

piece from the field is used.


Procedure:

While in <

Parts counting >

press

,

Using

or

choose a product and confirm it by pressing

.


Notice:
The selected product has to have declared unit mass (single piece mass).