Switch on continuous transmission in basic unit, Switch off continuous transmission in basic unit, Switch on continuous transmission in current unit – RADWAG WLY Scales User Manual
Page 189: Syntax: c1 cr lf, Possible answers, Frame format, Syntax: c0 cr lf, Possible answers answers, Syntax: cu1 cr lf

189
34.4.10. Switch on continuous transmission in basic unit
Syntax: C1 CR LF
Possible answers:
C1_I CR LF
- command comprehended but cannot be executed
C1_A CR LF
MASS FRAME
- command comprehended and in progress
- mass value in basic unit is returned
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
I
space
stability
space
sign
mass
space
unit
CR
LF
34.4.11. Switch off continuous transmission in basic unit
Syntax: C0 CR LF
Possible answers answers:
C0_I CR LF
- command comprehended but cannot be executed
C0_A CR LF
- command comprehended and executed
34.4.12. Switch on continuous transmission in current unit
Syntax: CU1 CR LF
Possible answers:
CU1_I CR LF
- command comprehended but cannot be executed
CU1_A CR LF
MASS FRAME
- command comprehended and in progress
- mass value in current unit is returned
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
U
I
stability
space
sign
mass
space
unit
CR
LF