beautypg.com

Switch on continuous transmission in basic unit, Switch off continuous transmission in basic unit, Switch on continuous transmission in current unit – RADWAG WLY Scales User Manual

Page 189: Syntax: c1 cr lf, Possible answers, Frame format, Syntax: c0 cr lf, Possible answers answers, Syntax: cu1 cr lf

background image

189

34.4.10. Switch on continuous transmission in basic unit

Syntax: C1 CR LF

Possible answers:

C1_I CR LF

- command comprehended but cannot be executed

C1_A CR LF
MASS FRAME

- command comprehended and in progress

- mass value in basic unit is returned


Frame format:

1

2

3

4

5

6

7-15

16

17

18

19

20

21

S

I

space

stability

space

sign

mass

space

unit

CR

LF

34.4.11. Switch off continuous transmission in basic unit

Syntax: C0 CR LF

Possible answers answers:

C0_I CR LF

- command comprehended but cannot be executed

C0_A CR LF

- command comprehended and executed

34.4.12. Switch on continuous transmission in current unit

Syntax: CU1 CR LF

Possible answers:

CU1_I CR LF

- command comprehended but cannot be executed

CU1_A CR LF
MASS FRAME

- command comprehended and in progress

- mass value in current unit is returned


Frame format:

1

2

3

4

5

6

7-15

16

17

18

19

20

21

S

U

I

stability

space

sign

mass

space

unit

CR

LF

This manual is related to the following products: