16.3.1. Port for barcode scanner


Procedure:

Enter <

Devices>

according to ch.16 of the manual, choose

Barcode reader

/

Port”

and then set the appropriate value.

Barcode scanners can be connected to:

RS232 (1),

RS232 (2).

16.3.2. Prefix / Suffix

Users can edit a prefix <

Prefix>

or / and suffix <

Suffix>

in order

to adjust the program to accept transmission frames from the scanner.


Notice:
A special protocol is required in order the code be received by RADWAG
equipment. It is required to program an appropriate prefix and suffix.
Prefix – one byte 01 hexadecimally, suffix one byte 0D hexadecimally.
The detailed description of cooperation scale – barcode scanner can
be found in APPENDIX F in this manual.

Procedure:

Enter <

Barcode Scanner>

according to ch.16.3 of the manual,

Chose parameter <

Prefix>

and then enter, using the screen

keyboard, a required value (hexadecimal) and confirm it by pressing

.

Chose parameter <

Suffix>

and then enter, using the screen

keyboard, a required value (hexadecimal) and confirm it by pressing

.

16.3.3. Field selection

This option is connected with selecting data which the program will search

after reading a barcode.