beautypg.com

Send the result immediately in current unit, Frame format, Syntax: sui cr lf – RADWAG WLY Scales User Manual

Page 188: Possible answers

background image

188

SU_A CR LF
SU_E CR LF

- command accepted and in progress

- timeout while waiting for stable results

SU_I CR LF

- command comprehended but cannot be executed

SU_A CR LF
MASS FRAME

- command accepted and in progress

- mass value in current unit is returned


Frame format:

1

2

3

4

5

6

7-15

16

17

18

19

20

21

S

U

space

stability

space

sign

mass

space

unit

CR

LF

Example:

S U CR LF

– computer command

S U _ A CR LF

- command accepted and in progress

S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF

- command done, mass

value in current unit is returned.

34.4.9. Send the result immediately in current unit

Syntax: SUI CR LF

Possible answers:

SUI_I CR LF

- command comprehended but cannot be executed

MASS FRAME

- mass value in current unit is returned immediately


Frame format:

1

2

3

4

5

6

7-15

16

17

18

19

20

21

S

U

I

stability

space

sign

mass

space

unit

CR

LF

Example:

S U I CR LF

– computer command

S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF

- command executed and mass

returned

This manual is related to the following products: