RADWAG WLY Scales User Manual
Page 131

131
Mass
Product nominal mass
Tare
Product tare value (set automaticallz on
selecting a product from the database)
CPG
-
CPG mode
Control type: Non-destructive Average Tare,
Non-destructive Empty-Full, Destructive Full-
Empty, Destructive Empty-Full
Lot size
Declaring quantity of a controlled batch
Charge
Measuring series to be controlled: Non-
destructive Empty-Full, Destructive Full-Empty,
Destructive Empty-Full
Unit
Product’s weighing unit: [g] or [ml]
Density
Product density (range of entered values has
to contain between 0,1g/cm
3
and 5g/cm
3
)
Quantity of
packages
Declaring quantity of packages to be controlled
for determining average tare (for control type
“non-destructive average tare”)
Internal control
Submenu for defining internal criteria of a
control (see table below)
•
List of data for internal criteria
Internal control
Enabling
/ disabling
internal control
criteria
Sample quantity
Quantity of a product sample
Erro value [- T]
Value of negative permissible error –T, set in
a weighing unit determined for the product.
Measurements below the Qn-T value will
be comprehended as defective.
Erro value [+ T]
Value of positive permissible error +T set in
a weighing unit determined for the product.
Measurements above the Qn+T value will
be comprehended as defective.
Quantity of disqualifying
samples [Qn – 2T]
Number of recorded negative errors -2T in
a tested sample that disqualifies a control
Quantity of disqualifying
samples [Qn + 2T]
Number of recorded positive errors +2T in
a tested sample that disqualifies a control