Local parameters
Select formulation
Start formula making process
Stop formula making process
Select formulation ingredient from the list
Select previous formulation ingredient to to be processed
Select net formulation ingredient to be processed
Set “mass from hand” – mass of an ingredient supplied in ready
packaged and with determined mass
Edit lot number of an ingredient and/or sample of a formulation
ingredient
29.2. Local setting of a working mode
Local settings for the working mode <
Formulation >
are accessible
on pressing a hot key <
Local parameters>:
Ask for multiplier
Causes setting a question for formulation
multiplier, i.e. the number which is used to
multiply the values of mass of all formulation
components.
Ask for number
of cycles
Causes setting a question on number of
formulation cycles, i.e. number of repetitions of
the same formulation.
Confirm batching
ingredients manually
Causes enabling manual confirmation by
pressing “Enter/Print” key on terminal’s keyboard
for each weighing process.
Automatic tare
Enable automatic tarring of mass at the moment
of process start and mass of each following
ingredient after the sampling process.
Ingredient control
Enable control mode of an ingredient being
part of a formulation. The “Ingredient control”
mode orders giving correct ingredient code
before its weighing process.