beautypg.com

Dosing process structure, Local setting of a working mode – RADWAG WLY Scales User Manual

Page 107

background image

107

28.2. Dosing process structure

All activities related to the dosing process can be carried out from
the level of scale. Each dosing process <

>

contains:

Its name <

>

,

code <

>

,

assigned weighing platforms <

>

defined on a terminal.

Each of the weighing platforms can have its specific dosing process. The

database of dosing processes has the same searching mechanisms for a

specific record in a database as other terminal databases.

A dosing process can be searched in a database by:

the name of a dosing process <

>

the code of a dosing process <

>

Processes carried out on different weighing platforms operated by the same

terminal, can also be dependant on each other, e.g.: dosing on weighing

platform 2 can be initiated only on completing dosing on weighing platform 1,

which is confirmed by a signal from appropriate sensor.

28.3. Local setting of a working mode

Local settings for the working mode <

Dosing>

are accessible

on pressing a hot key <

Local parameters>:

Ask for multiplier

Causes setting a question for dosing

process multiplier, i.e. the number which is

used to multiply the values of mass of all

components.

Ask for number of cycles

Causes setting a question on number of

dosing process cycles, i.e. number of

repetitions of the same process.

Confirm batching

ingredients manually

Causes enabling manual confirmation by

pressing “Enter/Print” key on terminal’s

keyboard for each weighing process.

No. of weighings for

calculating the correction

Determines the number of recent

measurements which are analyzed for

calculating automatic correction in dosing

This manual is related to the following products: