beautypg.com

Send the stable result in current unit – RADWAG WLY Scales User Manual

Page 187

background image

187

Example:

S I CR LF

– computer command

S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF

- command done, mass value

in basic unit is returned immediately.

34.4.7. Send immediate results from all platforms in basic units

Syntax: SIA CR LF

Possible answers:

SIA_I CR LF

- command comprehended but cannot be executed at the moment

MASS FRAME „P1” CR LF
MASS FRAME „P2” CR LF

- mass values are immediately returned from all

platforms in basic units

Frame format with mass from subsequent platforms as indicator reply:

1

2

3

4

5

6

7-15

16

17

18

19

20

21

P

n

space

stability

space

sign

mass

space

unit

CR

LF

n

- weighing platform number

mass

- 9 characters justified to the right

unit

- 3 characters justified to the left

Example:

Let us assume that both platforms are connected to scale.
S I A CR LF

– computer command

P 1 _ ? _ _ _ _ _ _ 1 1 8 . 5 _ g _ _ CR LF
P 2 _ _ _ _ _ _ _ _ _ 3 6 . 2 _ k g _ CR LF

- command done, mass values

from both platforms are returned in basic units

34.4.8. Send the stable result in current unit

Syntax: SU CR LF

Possible answers: