Filter, Autozero – RADWAG WLY Scales User Manual

Page 42

background image

42

Procedure:

Enter <

Weighing>

according to ch. 14 of the manual, select

<

Median Filter>

and then set an appropriate value.


Accessible settings:

None

- median filter is off

0.5, 1, 1.5, 2, 2.5

- filtering level to choose

14.2. Filter

This filter is intended to suppress continuous mechanical vibrations at the

cost of stabilization time.

Procedure:

Enter <

Weighing>

according to ch. 14 of the manual, select

<

Filter>

and then set an appropriate value.


Accessible settings:

None, V.Fast, Fast, Average, Slow.

Notice:
The higher filtering level the longer stabilization time.

14.3. Autozero

The autozero function has been implemented in order to assure precise

indications. This function controls and corrects „0” indication.

While the function is active it compares the results continuously with

constant frequency. If two sequentional results differ less than the declared

value of autozero range, so the scale will be automatically zeroed and the

pictograms

and

will be displayed.

If AUTOZERO is disabled zero is not corrected automatically. However,

in particular cases, this function can disrupt the measurement process e.g.

slow pouring of liquid or powder on the weighing pan. In this case, it is

advisable to disable the autozero function.