Select the capacitance range – GW Instek GDM-8300 series User Manual User Manual

Page 47

background image

OPERATION

47

Select the Capacitance Range

The capacitance range can be set automatically or manually.

Auto Range

To turn the automatic range selection On/Off,
press the AUTO key.

Manual Range

Press the Up or the Down key to select the
range. The AUTO indicator turns Off
automatically. If the appropriate range is
unknown, select the highest range.

Selectable
Capacitance
Ranges

Range

Resolution Full scale

5nF

1pF

5.100nF

50nF

10pF

51.00nF

500nF

100pF

510.0nF

5μF

1nF

5.100μF

50μF

10nF

51.00μF

Note

For further details, please see the specifications on
page 152.

Note

The refresh rate settings and the EXT trigger
cannot be used in the capacitance mode.