beautypg.com

Trigger commands, Read, Val1 – GW Instek GDM-8300 series User Manual User Manual

Page 134: Val2, Trigger:source

background image

GDM-834X Series User Manual

134

TRIGger Commands

READ?

Returns 1

st

and 2

nd

display value.

VAL1?

Returns the 1

st

display reading

Example: SAMP:COUN 100

VAL1?

>+0.333E-4,V DC
>+0.389E-4,V DC
> etc, for 100 counts.
Queries 100 counts of stored samples from the 1

st

display.

VAL2?

Returns the 2

nd

display reading.

Example: SAMP:COUN 100

VAL2?

>+0.345E-4,V DC
>+0.391E-4,V DC
> etc, for 100 counts.
Queries 100 counts of stored samples from the 2

nd

display.

TRIGger:SOURce

Selects the trigger source.
Parameter: INT | EXT
Example: TRIG:SOUR INT
Sets the trigger source as internal.

TRIGger:SOURce?

Returns current trigger source.