APPENDIX

153

Capacitance

Range

Resolution

Full Scale Test Current

Accuracy

(1 year 23˚C ±5˚C) [1]

5nF: 0.5~1nF [2]

0.001nF

5.100

8.3

μ

A

2.0%+20

5nF: 1~5nF [2]

2.0%+10

50nF: 5~10nF [2]

0.01nF

51.00

8.3

μ

A

2.0%+30

50nF: 10~50nF [2]

2.0%+10

500nF

0.1nF

510.0

83

μ

A

2.0%+4

5

μ

F

1nF

5.100

0.56mA

50

μ

F

10nF

51.00

0.83mA

[1] For the 5nF ~ 50

μF range, make sure that the input is greater than 10% of the

range.
[2] Need to use the REL function.
* Input protection of 500V peak on all ranges.

Frequency

Measurement Range

Accuracy (1 year 23˚C ±5˚C)

10Hz ~ 500Hz

0.01%+5

500Hz ~ 500kHz

0.01%+3

* AC + DC measurements do not allow frequency measurements.
* Input protection of 1000V peak on all ranges.

Voltage Measurement Sensitivity

Range

Minimum Sensitivity (RMS sine wave)

10~100kHz

100K~500kHz

500mV

35mV

200mV

5V

0.25V

0.5V

50V

2.5V

5V

500V

25V

uncal

750V

50V

uncal

Current Measurement Sensitivity

Range

Minimum Sensitivity (RMS sine wave)

30~20kHz

500

μA

35

μA

5mA

0.25mA

50mA

2.5mA

500mA

25mA

5 A

0.25A(<2kHz)

10 A

2.5A(<2kHz)