GDM-834X Series User Manual

118

CONFigure:PERiod

Sets measurement to Period on the first display and specifies the
range.
Parameter: [None] | [Range( | MIN | MAX | DEF)]
Example: CONF:PER
Sets the DMM to period measurement using the previous range.

CONFigure:CONTinuity

Sets measurement to Continuity on the first display.
Parameter: None

CONFigure:DIODe

Sets measurement to Diode on the first display.
Parameter: None

CONFigure:TEMPerature:TCOuple

Sets measurement to Temperature thermocouple (T-CUP) on the
first display.
Parameter: [None] | [Type(J | K | T)]
Example: CONF:TEMP:TCO J
Sets the measurement mode to TCO with a type J sensor.

CONFigure:CAPacitance

Sets measurement to Capacitance on the first display.
Parameter: [None] | [Range( | MIN | MAX | DEF)]
Example: CONF:CAP 5E-5
Sets the measurement mode to Capacitance with a 50μF Range.

CONFigure:FUNCtion?

Returns the current function on the first display.
Return parameter: VOLT, VOLT:AC,VOLT:DCAC, CURR,
CURR:AC,CURR:DCAC, RES, FREQ, PER, TEMP, DIOD, CONT,
CAP