beautypg.com

Resistance, Diode, Continuity – GW Instek GDM-8300 series User Manual User Manual

Page 152

background image

GDM-834X Series User Manual

152

Resistance

Resistance

Resolution

Full Scale

Test Current

Accuracy

(1 year 23˚C

±5˚C)[2]

500Ω

10mΩ

510.00

0.83mA

0.1%+5 [1]

5kΩ

100mΩ

5.1000

0.83mA

0.1%+3 [1]

50kΩ

1Ω

51.000

83

μ

A

0.1%+3

500kΩ

10Ω

510.00

8.3

μ

A

0.1%+3

5MΩ

100Ω

5.1000

830nA

0.1%+3

50MΩ

1KΩ

51.000

560nA//10MΩ

0.3%+3

[1] Using the REL function. If you don't use the REL function then increase the
error by 0.2Ω.
[2] When measuring resistances greater than 500kΩ, please use shielded test leads
to eliminate the noise interference that may be induced by standard test leads.
* Open circuit voltage approximates 6V max on 500~5MΩ range, approximates
5.5V max on 50MΩ range.
* Input protection of 500V peak on all ranges.

Diode

Range

Resolution

Full Scale

Test Current

Accuracy

(1 year 23˚C ±5˚C)

5V

100

μ

V

5.1000

0.83mA

0.05%+5

* Input protection of 500V peak. *Open circuit voltage approximates 6V.

Continuity

Range

Resolution

Full Scale Test Current

Accuracy

(1 year 23˚C ±5˚C)

5000.0Ω

100mΩ

5100.0

0.83mA

0.1%+5