beautypg.com

Select the voltage range – GW Instek GDM-8300 series User Manual User Manual

Page 35

background image

OPERATION

35

Select the Voltage Range

The voltage range can be set automatically or manually.

Auto Range

To turn the automatic range selection On/Off,
press the AUTO key.

Manual Range

Press the Up or the Down key to select the
range. The AUTO indicator turns Off
automatically. If the appropriate range is
unknown, select the highest range.

Selectable Voltage
Ranges

Range

Resolution

Full scale

500mV

10μV

510.00mV

5V

0.1mV

5.1000V

50V

1mV

51.000V

500V

10mV

510.00V

750V (AC)

100mV

765.0V

1000V (DC)

100mV

1020.0V

Note

For further details, please see the specifications on
page 149.

Note

DC voltages with AC components cannot be
accurately measured if the DC+AC component
exceeds the ADC dynamic range for the selected
DC range. Any voltage exceeding the ADC dynamic
range will be clipped at the upper/lower range
limit. Under these conditions the range that is
chosen with the Auto range function may be too
small.