beautypg.com

Sense commands – GW Instek GDM-8300 series User Manual User Manual

Page 127

background image

COMMAND OVERVIEW

127

SENSe Commands

[SENSe:]TEMPerature:TCOuple:TYPE

Sets thermocouple type.
Parameter: Type(J | K | T)
Example: SENS:TEMP:TCO:TYPE J
Sets the thermocouple to type J.

[SENSe:]TEMPerature:TCOuple:TYPE?

Returns the thermocouple type.
Return parameter: J, K, T

[SENSe:]TEMPerature:RJUNction:SIMulated

Set temperature simulation value.
Parameter: (0.00 ~ 50.00)
Example: SENS:TEMP:RJUN:SIM 25.00
Sets the thermocouple junction temperature to 25˚C.

[SENSe:]TEMPerature:RJUNction:SIMulated?

Returns temperature simulation value.
Return parameter: (+0000~+5000) ,where +0000=0.00˚C,
+5000=50.00˚C

[SENSe:]DETector:RATE

Sets the detection rate (sample rate)
Parameter: RATE(S | M | F)
Example: SENS:DET:RATE S
Sets the rate to slow (S).

[SENSe:]DETector:RATE?

Returns the sample rate.
Return parameter: SLOW, MID, FAST