COMMAND OVERVIEW

109

[MIN] (Optional
parameter)

For commands, this will set the
setting to the lowest value. This
parameter can be used in place of
any numerical parameter where
indicated.
For queries, it will return the
lowest possible value allowed for
the particular setting.

[MAX] (Optional
parameter)

For commands, this will set the
setting to the highest value. This
parameter can be used in place of
any numerical parameter where
indicated.
For queries, it will return the
highest possible value allowed
for the particular setting.

Automatic
parameter range
selection

The GDM-8342/8341 automatically sets the
command parameter to the next available value.

Example

conf:volt:dc 1
This will set the measurement
item to DC Voltage and the range
to 5V. There is no 1V range so the
DMM selects the next available
range, 5V.

Message
Terminator (EOL)

Remote
Command

Marks the end of a command
line. The following messages are
in accordance with IEEE488.2
standard.

LF, CR, CR+LF

The most
common EOL
character is
CR+LF