beautypg.com

Ac voltage, acv+dcv, Ac coupled), Ac current, aci+dci – GW Instek GDM-8300 series User Manual User Manual

Page 151

background image

APPENDIX

151

AC Voltage, ACV+DCV

[3]

(AC Coupled)

Range Resolution

Full

Scale

Accuracy (1 year 23˚C ±5˚C) [1]

30-50Hz

50-10kHz

10K-30kHz 30K-100kHz

500mV

10

μ

V

510.00

1.00%+40

0.50%+40

2.00%+60

3.00%+120

5V

100

μ

V

5.1000

1.00%+20

0.35%+15

1.00%+20

3.00%+50

50V

1mV

51.000

1.00%+20

0.35%+15

1.00%+20

3.00%+50

500V

10mV

510.00

x

0.5%+15

1.00%+20[2] 3.00%+50[2]

750V

100mV

765.0

x

0.5%+15

x

x

[1]Specifications are for sine wave inputs that are greater than 5% range.
[2]Input voltage <300Vrms.
[3]The accuracy of ACV+DCV is equal to ACV’s with 10 more digits added.
* The specifications are guaranteed to an input of 750V. A beeping alarm will go off
when the input value is higher than 750V.
* Input protection of 1000V peak on all ranges.
* AC-coupled true RMS – measures the AC component of the input with up to
400Vdc of bias on any range.

AC Current, ACI+DCI

[3]

(AC Coupled)

Range

Resolu-

tion

Full

Scale

Accuracy (1 year 23˚C ±5˚C) [1]

Burden

Voltage

30-50Hz

50-2kHz

2K-5kHz

5K-20kHz

500

μ

A

10nA

510.00

1.50%+50 0.50%+40 1.50%+50 3.00%+75 0.06V max

5mA

100nA 5.1000

1.50%+40 0.50%+20 1.50%+40 3.00%+60

0.6V max

50mA

1

μ

A

51.000

1.50%+40 0.50%+20 1.50%+40 3.00%+60 0.14V max

500mA 10

μ

A

510.00

1.50%+40 0.50%+20 1.50%+40 3.00%+60[2] 1.4V max

5A

100

μ

A 5.1000

2.0%+40 0.50%+30

x

x

0.5V max

10A

1mA

12.000

2.0%+40 0.50%+30

x

x

0.8V max

[1] The 500

μ

A range requires an input of >35

μ

A to meet specifications. The

5mA~10A ranges need more than 5% of full scale range to meet specifications.
[2] Input current (5k ~ 20kHz)<330mArms.
[3]The accuracy of ACI+DCI is equal to ACI’s with 10 more digits added.
* The specifications are guaranteed to 10A. A beeping alarm will go off when the
input current being measured is higher than 10A.