CONFigure Commands

CONFigure:VOLTage:DC

Sets measurement to DC Voltage on the first display and
specifies the range.
Parameter: [None] | [Range( | MIN | MAX | DEF)]
Example: CONF:VOLT:DC 5
Sets the voltage range to 5 volts.

CONFigure:VOLTage:AC

Sets measurement to AC Voltage on the first display and
specifies the range.
Parameter: [None] | [Range( | MIN | MAX | DEF)]
Example: CONF:VOLT:AC
Sets the AC range to auto range.

CONFigure:VOLTage:DCAC

Sets measurement to DC+AC Voltage on the first display and
specifies the range.
Parameter: [None] | [Range( | MIN | MAX | DEF)]
Example: CONF:VOLT:DCAC
Sets the DC+AC voltage range to auto range.

CONFigure:CURRent:DC

Sets measurement to DC Current on the first display and
specifies the range.
Parameter: [None] | [Range( | MIN | MAX | DEF)]
Example: CONF:CURR:DC 50e-3
Sets the DC current range to 50mA.