beautypg.com

Specifications, General specifications – GW Instek GDM-8300 series User Manual User Manual

Page 149

background image

APPENDIX

149

Specifications

The specifications apply when the DMM is warmed up for at least
30 minutes and operates in slow rate.
Below are the basic conditions required to operate the DMM within
specifications:

Calibration: Yearly

Operating Temperature Specification: 18~28˚C (64.4~82.4˚F)

Relative Humidity: 80% (Non condensing)

Accuracy: ± (% of Reading + Digits)

AC measurements are based on a 50% duty cycle.

The power supply cable must be grounded to ensure accuracy.

All specifications are applicable to the main (1

st

) display only.

General Specifications

Specification Conditions:
Temperature: 23˚C±5˚C
Humidity: <80%RH, 75%RH for resistance measurement readings greater than
10MΩ.
Operating Environment: (0~50˚C)
Temperature Range: 0~35˚C, Relative Humidity: <80%RH;
>35˚C, Relative Humidity: <70%RH
Indoor use only
Altitude: 2000 meters
Pollution degree 2
Storage Conditions (-10~70˚C)
Temperature Range: 0~35˚C, Relative Humidity: <90%RH;
>35˚C, Relative Humidity: <80%RH
General:
Power Consumption: Max 15VA
Dimensions: 265 mm (W) X 107 mm (H) X 302 mm (D)
Weight: Approximately 2.9 kg