41

Select the Current Range

The current range can be set automatically or manually.

Auto Range

To turn the automatic range selection On/Off,
press the AUTO key.

Manual Range

Press the Up or the Down key to select the
range. The AUTO indicator turns Off
automatically. If the appropriate range is
unknown, select the highest range.

Selectable
Current Ranges

Range

Resolution Full scale

500µA

10nA

510.00μA

5mA

100nA

5.1000mA

50mA

1μA

51.000mA

500mA

10μA

510.00mA

5A

100μA

5.1000A

10A

1mA

12.000A

Note

For further details, please see the specifications on
page 149.

Note

DC currents with AC components cannot be
accurately measured if the DC+AC component
exceed the ADC dynamic range for the selected DC
range. Any current exceeding the ADC dynamic
range will be clipped at the upper/lower range
limit. Under these conditions the range that is
chosen with the Auto range function may be too
small.