Displaying a Single Reading. In the sensor calibration program, it is possible to take a single reading and
display it in the Sensor Configuration [Calibration] window. First, click on the Channel Values tab. Then
click on the Take a Single Reading button on the tool bar, or choose Take a Single Reading from the
Calibration menu item.

Displayed readings are based on the most recent calibration. Changing the calibration parameters,
without calibrating the system, will not affect the channel values.

Channel Values Tab Selected

Channel Values, Simplified Block Diagram

The simplified block diagram (above) can be used to better understand the relationship of channel amplifiers and their
corresponding user interface columns (visible in the Channel Values Tab). These columns are represented in the tab
figure, and in the following table.

Column

Display Description

Bridge (mV)

The input value from the bridge. The value is in millivolts.

Input Gain

The Gain setting of WBK16’s Input Amplifier. Any one of the following four settings is possible:

x1, x10, x100, or x1000.

Offset Null (Volts)

The Voltage summed into WBK16s Scaling Amplifier. The voltage is to compensate for any

offset that is present in the sensor’s output. The Offset Null voltage is in the range of
-3 to +3 volts.

Scaling Gain

The Gain setting of WBK16’s Scaling Amplifier. Any one of the following 13 gain settings can

be used: 1.0, 1.28, 1.65, 2.11, 2.71, 3.48, 4.47, 5.47, 7.37, 9.46, 12.14, 15.58, or 20.0.

A/D In (Volts)

The digital voltage value that is received as input by WaveView software. This value is also

referred to as sensor output voltage.

Units Scaling

The Multiplier value [used by the software] for converting sensor output voltage into User Units.

Units Offset

The Offset value that is added to “Units Scaling” for fine adjustment of what will be the final

reading (Bridge Units).

Bridge (Units)

The Reading (in User Units, for example: lbs, psi, kg) that results from converting the initial

sensor reading (Bridge mV).

WBK16, pg. 28

949794

WBK16, Strain-Gage Module