WBK61 and WBK62, High-Voltage Adapters
988296
WBK61 and WBK62, pg. 5
WBK61 and WBK62 - Specifications
Name/Function:
WBK61, High-Voltage Adapter with Probes, 200:1 Voltage Divider
WBK62, High-Voltage Adapter with Probes, 20:1 Voltage Divider
Number of Channels: 1
Dimensions: 83 mm × 61 mm × 19.5 mm (3.25” × 2.375” × 0.77”)
Cables: 60” leads with detachable probe tips and alligator clips
Output Connector: BNC female
Voltage Divider:
WBK61: 200:1 fixed
WBK62: 20:1 fixed
Maximum Voltage
WBK61: 1000 Vpeak (on either input reference to earth ground)
WBK62: 100 Vpeak (on either input reference to earth ground)
Maximum Differential Voltage:
WBK61: 2000 Vpeak (if neither input exceeds 1000 Vp rating to earth ground)
WBK62: 200 Vpeak (if neither input exceeds 100 Vp rating to earth ground)
Frequency Characteristics: approximates a single-pole frequency response
-3 dB Bandwidth: 200 kHz minimum
WBK61, Effective Voltage Ranges
WBK62, Effective Voltage Ranges
Range
WaveBook
WBK10A
Range
WaveBook
WBK10A
Bipolar
±2000V
±1000V
±400V
±200V
±100V*
±40V*
±20V*
±2000V
±1000V
±400V
±200V
±100V*
±40V*
±20V*
±10V*
Bipolar
±200V
±100V
±40V
±20V
±10V*
±4V*
±2V*
±200V
±100V
±40V
±20V
±10V*
±4V*
±2V*
±1V*
Unipolar
Not Available
Unipolar
Not Available
0 to 2000V
0 to 1000V
0 to 400V
0 to 200V
0 to 100V*
0 to 40V*
0 to 20V*
0 to 200V
0 to 100V
0 to 40V
0 to 20V
0 to 10V*
0 to 4V*
0 to 2V*
*Available through the WBK11A, WBK12A, or WBK13A option cards.
Measurement Errors:
The following values include total system error, i.e., they include errors from WaveBook, WBK10A,
WBK11A, WBK12A, and WBK13A. The value for gain error does not include offset error.
Gain Error:
0.1% FS (unipolar)
0.2% FS (bipolar)
Offset Error:
0.1% FS (unipolar)
0.2% FS (bipolar)