Calibration Parameters Section of Window

Two Views Obtained by Scroll Bar Movement

Column

Description

Exc. Volts

Used to change the excitation voltage. Choose between 10.0, 5.0, 2.0, 1.0, .5, and “Off.”

Changing the excitation voltage on any channel between one and four will change the value
on all four lower channels. Likewise, changing the excitation voltage on any channel five
through eight will change the value on all four upper channels. Changing the Excitation
Voltage here will also change the Source Level column in the WaveView Configuration
main window.

Gage Factor

Used for calibrating strain gages with the Name Plate calibration method. To change this

value, select the box and enter a number greater than 0 and less than 1000. The fill option
is available for this column (see Units).

Sensitivity
(mV/V)

This column is used for calibrating a load cell or transducer using the Name Plate

calibration method. To change this value, select the box and enter a number greater
than 0 and less than 1000. The fill option is available for this column (see Units).

Shunt Resistor
(Ohm)

This column is used for calibrating any sensor using the Shunt calibration method. The

value must equal the value of the shunt resistor in ohms. To change the value, select the
box and enter a number greater than 0 and less than 1000000. The shunt value must not
exceed the value entered as the maximum load. The fill option is available for this column
(see Units).

Bridge Resistor
(Ohm)

Used for calibrating any sensor using the Shunt calibration method. The value refers to the

bridge arm that is shunted during shunt calibration. To change the value, select the box and
enter a number from 60 to 1000. The fill option is available for this column (see Units).

Full Rated Load
(Units)

This column is used for calibrating a load cell or transducer using the Name Plate

calibration method. To change this value, select the box and enter a number greater than 0
and less than 100000. The full-rated load must be greater than the value entered for the
maximum applied load. The fill option is available for this column (see Units).

Max Applied Load
(Units)

Used for calibrating any sensor using any calibration method. To change the value, select

the box and enter a number greater than 0 and less than 1000000. This value must be
greater than the quiescent/tare value. The fill option is available for this column (see Units).

Quiescent/Tare
(Units)

This column is used for calibrating any sensor using any calibration method. The value

entered is the value of the quiescent load on the sensor. To change the value, select the
box and enter a number between –1000000 and 1000000. This value must be less than the
maximum applied load value. The fill option is available for this column (see Units).

Point 1
(mV)

This column is used for calibrating any sensor using the Shunt, or 2-Point Automatic

calibration method. The number must equal the input value, in mV, of the first point in the
calibration. To change the value, select the box and enter a number between -10000 and
10000. The fill option is available for this column (see Units).

Point 1
(Units)

This column is used for calibrating any sensor using the Shunt, 2-Point Automatic, or

2-Point Manual calibration method. The number must equal the value, in the selected
units, of the first point in the calibration. To change the value, select the box and enter a
number between -1000000 and 1000000. The fill option is available for this column (see
Units).

Point 2
(mV)

Used for calibrating any sensor using the 2-Point Automatic calibration method. The

number must equal the input value, in mV, of the second point in the calibration. To change
the value, select the box and enter a number between -10000 and 10000. The fill option is
available for this column (see Units).

Point 2
(Units)

This column is used for calibrating any sensor using the 2-Point Automatic, or 2-Point

Manual calibration method. The number must equal the value, in the selected units, of
the second point in the calibration. To change the value, select the box and enter a number
between -1000000 and 1000000. The fill option is available for this column (see Units).

WBK16, Strain-Gage Module