beautypg.com

Measurement Computing WBK Options User Manual

Page 49

background image

Half-Bridge Configurations

The half-bridge has two strain-variable elements and requires two internal bridge completion resistors (BCRs).
Any bridge resistance from 60 to 1000 ohms can be accommodated for either the internal or external bridge.

Half-Bridge (+), Any Resistance from 60 to 1000 Ohms,

B,D, or F Shunt

IA

External
Bridge

DB-9
Input

Internal Bridge Completion

RA

RB

RC

RD

RE

RF

RG

RH

1

2

3

-Sense

+Sense
+Excitation

-Excitation

JN01

Switches
accessed
through
software

1

8

9

7

5

4

3

6

2

H
G
F
E
D
C
B
A

Shunt

BCR

BCR

Shunt

Shunt

Shunt

In this connection, excitation voltage is
regulated at the connector.

This configuration should only be used
for short cable lengths
.

Output polarity can be altered by
selecting the software invert function.
The B, D, or F shunt resistor may be
software selected. Output polarity during
shunt calibration will be automatically
corrected by software.

IA

External
Bridge

DB-9
Input

Internal Bridge Completion

RA

RB

RC

RD

RE

RF

RG

RH

1

2

3

-Sense

+Sense
+Excitation

-Excitation

JN01

Switches
accessed
through
software

Half-Bridge (+), with Remote Sense

1

8

9

7

5

4

3

6

2

In this connection, excitation voltage is
regulated at the strain gage.

This is the preferred connection for
longer cables
.

3-Wire TC Half-Bridge, Software Invert & B, D, F Shunt Available

IA

External
Bridge

DB-9
Input

Internal Bridge Completion

RA

RB

RC

RD

RE

RF

RG

RH

1

2

3

-Sense

+Sense
+Excitation

-Excitation

JN01

Switches
accessed
through
software

1

8

9

7

5

4

3

6

2

This occasionally utilized connection can
be made as shown. Two resistors
normally reserved for quarter-bridge
completion must be used.

For compatibility with other
configurations, use of one of the above
two configurations is preferred over
this one
.

WBK16, Strain-Gage Module

949794

WBK16, pg. 9

This manual is related to the following products: