beautypg.com

Excitation connection, Caution – Measurement Computing WBK Options User Manual

Page 46

background image

Excitation Connection

Remote sense inputs are provided for the excitation regulators. The excitation voltage will be most accurate
at points where remote sense lines are connected, preferably at the bridge (this is often referred to as a
6-wire connection). Long cables will reduce the voltage at the bridge, due to current flow and wire
resistance, if remote sense is not used.

If the 6-wire approach is not used, the remote sense inputs must be jumpered to the excitation
outputs at the input connector
. Internal 1 M

Ω resistors are also connected where the jumpers would be

located to prevent circuit discontinuities. These 1 M

Ω resistors are not suitable for high-accuracy

excitation-voltage regulation.

3-wire quarter-bridge configurations do not benefit from external remote sense connections. The lead
resistance is actually a balanced part of the bridge. If the + remote sense input is connected to the + input
on a quarter-bridge, the voltage is regulated across the bridge completion resistor. This results in a
constant-current linearized quarter-bridge; otherwise quarter-bridges are not perfectly linear.

Shunt-Calibration Resistors. StrainBooks and WBK16s each provide three physical locations for internal
shunt-calibration resistors for each channel. Each shunt resistor is switched in from the EXCITATION (-)
to the IN (+) of the Instrumentation Amp by a FET switch to create a repeatable bridge imbalance. Internal
resistance of the circuit is about 1 k

Ω; the exact amount is automatically accounted for in the software. The

software also allows selection of the three shunt resistors ( B, D, F ). An internal inversion stage insures
correct polarity during the shunt calibration process; which arm is shunted is therefore irrelevant. Header
positions Rb, Rd, Rf correspond to the software shunt resistor selections of B, D, F.

For any balanced bridge, a resistance value can be placed in parallel with one element to create a
predictable imbalance and output voltage. This shunt-resistance value can be calculated by the following
equation, where V

out

is the differential output voltage of the gage.

Example:

R

Shunt

= R

Bridge Arm

[ ( V

Excitation

/ 4 (V

out

)) - 0.5 ]

R

Shunt

= 350 [ ( 10 / 4(0.020)) - 0.5 ] = 43,575

CAUTION

Be careful to avoid component damage while the WBK16 enclosure is open. Always
remove bridge completion headers (adapter plugs) from the unit before soldering
resistors in the headers.

Configuring the Bridge Completion Resistor Modules. For each channel, the
board has a 2×8 resistor socket with rows designated A through H. The removable
adapter plugs are included for soldering in the resistors. Additional adapter plugs
are available for convenient changeover of alternate configurations. Resistor Ra is
located nearest the front panel.

o

Half-bridge completion resistors consist of Rg and Rh.

o

Quarter-bridge completion resistors consist of Ra, Rc, and Re.

o

Shunt resistors consist of Rb, Rd, and Rf.

WBK16, pg. 6

949794

WBK16, Strain-Gage Module

Inserting resistors directly into the socket makes an unreliable connection and
is not recommended. Remove the plug from the main board; then solder
resistors to the adapter plug as indicated. To avoid damaging the pin
alignment on the plug, solder with minimal heat. After soldering, the resistor
leads should be snipped off close to the support.

Solder resistor lead
into support fork.

Soldering Resistors to

Adapter Plug

This manual is related to the following products: