AMETEK Compact i/iX Series User Manual
Page 76
User Manual
Compact i/iX Series
AMETEK Programmable Power
California Instruments
76
4.2.10 CALIBRATION Menus
The measurement calibration menu can be used to perform routine calibration of the internal
measurement system. The recommended calibration interval is 12 months. To enter the calibration
screens, the calibration password must be entered first.
Note: Refer to chapter 6 for details on routine calibration procedures and equipment requirements. Do
not attempt calibration without consulting the user manual.
This menu also contains the LCD viewing angle adjustment.
ENTRY
DESCRIPTION
VIEW ANGLE
-10 to +10
LCD viewing angle adjustment.
CAL PWORD
V range
Calibration password required to access all
calibration screens. The calibration password
is the high voltage range value. [300 ] The
password can be entered using the keypad or
shuttle followed by the ENTER key.
Measurement Calibration Screens
MVOLT FS
± 9999
Calibration coefficient for full-scale voltage
measurement.
MCURR FS
± 9999
Calibration coefficient for full-scale current
measurement.
Output Calibration Screens
VOLT FS
± 9999
Calibration coefficient for voltage output.
PHASE OFST
0.0 - 360.0
Phase offset calibration for units equipped with
the
–LKS option. Determines phase offset
from
–LKM master clock.