10.6.2 Emergency Test

From the MIL704 main menu (Figure 10-2) scroll to the EMERGENCY entry using the up and down
cursor keys. Press the ENTER key to select the EMERGENCY screens. The screen will appear as
shown in Figure 10-4.

Figure 10-4: Emergency Menu

The EMERGENCY SCREEN has the following tests:

1

VOLTAGE

2

FREQUENCY

The above tests can be selected by scrolling to the highlighted selection using the up and down key
and the ENTER key to start the selected test.

Note: These tests are only required for revision D. See steady state voltage and frequency

tests for all other revisions.

VOLTAGE

This test will change the output voltage in the following sequence:

104V for 1 minute.

122V for 1 minute.

115V for 1 minute.

The

key (backspace) will terminate the test at any time.

FREQUENCY

This test will change the output frequency in the following sequence:

360Hz for 1 minute.

440Hz for 1 minute.

400Hz for 1 minute.

The

key (backspace) will terminate the test at any time.