AMETEK Compact i/iX Series User Manual

Page 142

background image

User Manual

Compact i/iX Series

AMETEK Programmable Power

California Instruments

142

10.7.2 Transient Test

From the MIL704 DC MENU scroll to the TRANSIENT DC entry using the up and down cursor keys.
Press the ENTER key to select the TRANSIENT screen.

The Transient Test has the following tests:

1

HIGH VOLTAGE

2

LOW VOLTAGE

HIGH VOLTAGE

This test will change the output voltage for the selected phase in the following sequence:

1. 28V System

50V for 12.5 msec.

Linearly reduce to29V in 70msec.

Stay at 29V for 92.5msec before returning to 28V.

2. 270V System

330V for 20 msec.

Linearly reduce to280V in 20msec.

Stay at 280V for 135msec before returning to 270V.

Prior to the test, a range change may take place if the power source is set for the low voltage range.

The

key (backspace) will terminate the test at any time.

Note: A range change will result in momentary loss of power to the EUT. If this is not acceptable, the
power source must be left in high range at all times.

LOW VOLTAGE

This test will change the output voltage for the selected phase in the following sequence:

1. 28V System

18V for 15 msec.

Linearly increase to 22V in 85msec.

Stay at 22V for 75msec before returning to 28V.

2. 270V System

200V for 10 msec.

Linearly increase to 250V in 30msec.

Stay at 250V for 135msec before returning to 270V.

The

key (backspace) will terminate the test at any time.