AMETEK Compact i/iX Series User Manual
Page 62
User Manual
Compact i/iX Series
AMETEK Programmable Power
California Instruments
62
ENTRY
DESCRIPTION
to switch between waveforms.
VOLT MODE
Selects the available output modes of operation. Available
modes are AC, DC (i/iX models) and ACDC (iX models only).
The shuttle can be used to select the desired output mode.
DC OFFSET
This parameter applies only when the power source is in ACDC
mode. The DC offset can only be set in percent of the AC RMS
voltage programmed. (Relative programming only). The
available dc offset range is ±20%.
Note: Changing the offset percentage setting will result in
temporary loss of the output voltage as the new offset is
recalculated and loaded. This may cause the EUT to reset or
turn off. To avoid this, set the desired offset percentage before
programming the AC voltage and turning on the output to the
EUT.
CURR
Sets the current limit value for the current detection system.
When the load current value exceeds the set current limit, a fault
condition is generated. The actual response of the AC Source to
a current limit fault is determined by the protection mode
selected in the OL MODE field. (CC = Constant Current, CV =
Constant Voltage).
OL MODE
Sets the current limit overload mode. The actual response of the
AC Source to a current limit fault is determined by this setting.
Available settings are CC for Constant Current mode or CV for
Constant Voltage mode. In CV mode, the AC source output will
trip off and stay off until re-engaged. In CC mode, the voltage will
be reduced until the current limit is no longer exceeded.